Fluke 5522A Operators Manual

Download or browse on-line these Operator's Manual for Fluke 5522A Operators Multimeter, Test Equipment.

Fluke 5522A Operators Manual Information:

This manual for Fluke 5522A Operators, given in the PDF format, is available for free online viewing and download without logging on. The guide contains 426 pages, and the size of the file at download is 13.13 Mb. The document type is Operator's Manual.

Download Manual

Summary of Contents:

[Page 1] Fluke 5522A Operators

January 2011 © 2011 Fluke Corporation. All rights reserved. Printed in USA. Specifications are subject to change without notice. All product names are trademarks of their respective companies. 5522A Multi-Product Calibrator Operators Ma...

[Page 2] Fluke 5522A Operators

LIMITED WARRANTY AND LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year and begins on the date of shipment. Parts, product repai...

[Page 3] Fluke 5522A Operators

OPERATOR SAFETY SUMMARY WARNING HIGH VOLTAGE is used in the operation of this equipment LETHAL VOLTAGE may be present on the terminals, observe all safety precautions! To avoid electrical shock hazard, the operator should not electricall...

[Page 4] Fluke 5522A Operators

...

[Page 5] Fluke 5522A Operators

i Table of Contents Chapter Title Page 1 Introduction and Specifications......................................................... 1-1 Introduction..............................................................................................

[Page 6] Fluke 5522A Operators

5522A Operators Manual ii AC Current (Non-Sine Wave) ....................................................................... 1-24 AC Current, Square Wave Characteristics .................................................... 1-25 AC Current, Tr...

[Page 7] Fluke 5522A Operators

Contents (continued) iii How to Set Resistance Output....................................................................... 4-33 How to Set Capacitance Output..................................................................... 4-34 How...

[Page 8] Fluke 5522A Operators

5522A Operators Manual iv How to Test the RS-232 UUT Port via IEEE-488 Port................................. 5-18 How to Change between Remote and Local Operation..................................... 5-20 Local State.............................

[Page 9] Fluke 5522A Operators

Contents (continued) v Introduction........................................................................................................ 6-3 Command Summary by Function .......................................................................

[Page 10] Fluke 5522A Operators

5522A Operators Manual vi The Leveled Sine Wave Function ................................................................. 9-15 Shortcuts for Setting the Frequency and Voltage.......................................... 9-16 The MORE OPTIONS Me...

[Page 11] Fluke 5522A Operators

Contents (continued) vii Trigger Signal Specifications (Edge Function) ............................................. 10-7 Trigger Signal Specifications (Square Wave Voltage Function)................... 10-7 TV Trigger Signal Specificatio...

[Page 12] Fluke 5522A Operators

5522A Operators Manual viii Leveled Sinewave Verification: Amplitude .................................................. 10-43 Leveled Sinewave Verification: Frequency .................................................. 10-44 Leveled Sinewave V...

[Page 13] Fluke 5522A Operators

Contents (continued) ix How to Set the Repeat Frequency ................................................................. 11-17 How to Set the Modulation Pattern ............................................................... 11-17 How t...

[Page 14] Fluke 5522A Operators

5522A Operators Manual x ...

[Page 15] Fluke 5522A Operators

xi List of Tables Table Title Page 1-1. Symbols.................................................................................................................. 1-4 2-1. Standard Equipment .....................................................

[Page 16] Fluke 5522A Operators

5522A Operators Manual xii 7-2. Replacement Current Fuses.................................................................................... 7-6 7-3. Verification Tests for DC Voltage (Normal) .....................................................

[Page 17] Fluke 5522A Operators

Contents (continued) xiii 10-17. SC1100 Option Leveled Sinewave Verification: Flatness ..................................... 10-46 10-18. SC1100 Option Marker Generator Verification..................................................... 10-...

[Page 18] Fluke 5522A Operators

5522A Operators Manual xiv ...

[Page 19] Fluke 5522A Operators

xv List of Figures Figure Title Page 1-1. 5522A Multi-Product Calibrator ............................................................................ 1-3 1-2. RS-232 Remote Connection..........................................................

[Page 20] Fluke 5522A Operators

5522A Operators Manual xvi 5-5. Testing the RS-232 UUT Port via RS-232 Host Port............................................. 5-17 5-6. Testing the RS-232 UUT Port via IEEE-488 Port ................................................. 5-19 5-7...

[Page 21] Fluke 5522A Operators

1-1 Chapter 1 Introduction and Specifications Title Page Introduction.......................................................................................................... 1-3 Safety Information ..............................................

[Page 22] Fluke 5522A Operators

5522A Operators Manual 1-2 ...

[Page 23] Fluke 5522A Operators

1-3 Introduction Warning If the 5522A Calibrator is operated in any way not specified by this manual or other documentation provided by Fluke, the protection provided by the Calibrator may be impaired. The 5522A Calibrator (hereafter “th...

[Page 24] Fluke 5522A Operators

5522A Operators Manual 1-4 • 10 MHz reference input and output. Use this to input a high-accuracy 10 MHz reference to transfer the frequency accuracy to the 5522A, or to synchronize one or more additional Calibrators to a master 5522A. • ...

[Page 25] Fluke 5522A Operators

Introduction and Specifications Overload Protection 1 1-5 • Read all safety Information before you use the Product. • Do not use the Product if it operates incorrectly. • Replace the mains power cord if the insulation is damaged or...

[Page 26] Fluke 5522A Operators

5522A Operators Manual 1-6 Remote Operation (RS-232) There are two rear-panel serial data RS-232 ports: SERIAL 1 FROM HOST, and SERIAL 2 TO UUT (see Figure 1-2). Each port is dedicated to serial data communications for operating and controlling...

[Page 27] Fluke 5522A Operators

Introduction and Specifications Where to Go from Here 1 1-7 can write your own programs using the IEEE-488 command set or run the optional Windows-based MET/CAL software. (See Chapter 6 for a discussion of the commands available for IEEE-488...

[Page 28] Fluke 5522A Operators

5522A Operators Manual 1-8 General Specifications The following tables list the 5522A specifications. All specifications are valid after allowing a warm-up period of 30 minutes, or twice the time the 5522A has been turned off. (For example, if t...

[Page 29] Fluke 5522A Operators

Introduction and Specifications Detailed Specifications 1 1-9 Detailed Specifications DC Voltage Absolute Uncertainty, tcal ±5 °C ±(ppm of output +μV) Stability Range 90 days 1 year 24 hours, ±1 °C ±(ppm, output +μV) Resoluti...

[Page 30] Fluke 5522A Operators

5522A Operators Manual 1-10 Noise Range Bandwidth 0.1 Hz to 10 Hz p-p Bandwidth 10 Hz to 10 kHz rms 0 to 329.999 μA 2 nA 20 nV 0 to 3.29999 mA 20 nA 200 nV 0 to 32.9999 mA 200 nA 2.0 μA 0 to 329.999 mA 2000 nA 20 μA 0 to 2.99...

[Page 31] Fluke 5522A Operators

Introduction and Specifications Detailed Specifications 1 1-11 Resistance Absolute Uncertainty, tcal ±5 °C ±(ppm of output +floor) [2] ppm of output Floor (Ω) Temp and temp since ohms zero cal Range [1] 90 days 1 year 12 hrs ±1...

[Page 32] Fluke 5522A Operators

5522A Operators Manual 1-12 AC Voltage (Sine Wave) Absolute Uncertainty, tcal ±5 °C ±(ppm of output + μV) Range Frequency 90 days 1 year Resolution Max Burden Max Distortion and Noise 10 Hz to 5 MHz Bandwidth ±(% output + floo...

[Page 33] Fluke 5522A Operators

Introduction and Specifications Detailed Specifications 1 1-13 AC Voltage (Sine Wave) (cont.) Absolute Uncertainty, tcal ±5 °C ±(% of output + μV) Range Frequency [1] 90 days 1 year Resolution Max Burden Max Distortion and N...

[Page 34] Fluke 5522A Operators

5522A Operators Manual 1-14 AC Current (Sine Wave) Absolute Uncertainty, tcal ±5 °C ±(% of output + μA) Range Frequency 90 days 1 year Compliance adder ±(μA/V) Max Distortion & Noise 10 Hz to 100 kHz BW ±(% of output + floo...

[Page 35] Fluke 5522A Operators

Introduction and Specifications Detailed Specifications 1 1-15 AC Current (Sine Wave) (cont.) Absolute Uncertainty, tcal ±5 °C ±(% of output + μA) Range Frequency 90 days 1 year Max Distortion & Noise 10 Hz to 100 kHz BW ±(...

[Page 36] Fluke 5522A Operators

5522A Operators Manual 1-16 Capacitance Absolute Uncertainty, tcal ±5 °C ±(% of output + floor) [1] [2] [3] Allowed Frequency or Charge-Discharge Rate Range 90 days 1 year Resolution Min and Max to Meet Specification Typical Ma...

[Page 37] Fluke 5522A Operators

Introduction and Specifications Detailed Specifications 1 1-17 Temperature Calibration (Thermocouple) Absolute Uncertainty Source/Measure tcal ±5 °C ± °C [3] Absolute Uncertainty Source/Measure tcal ±5 °C ± °C [3] TC Ty...

[Page 38] Fluke 5522A Operators

5522A Operators Manual 1-18 Temperature Calibration (RTD) Absolute Uncertainty tcal ±5 °C ± °C [2] Absolute Uncertainty tcal ±5 °C ± °C [2] RTD Type Range °C [1] 90 days 1 year RTD Type Range °C [1] 90 days 1 y...

[Page 39] Fluke 5522A Operators

Introduction and Specifications Detailed Specifications 1 1-19 AC Power (45 Hz to 65 Hz) Specification Summary, PF=1 Current Range 3.3 to 8.999 mA 9 to 32.999 mA 33 to 89.99 mA 90 to 329.99 mA Voltage Range Absolute Uncertainty, t...

[Page 40] Fluke 5522A Operators

5522A Operators Manual 1-20 Phase 1-Year Absolute Uncertainty, tcal ±5 °C, (Δ Φ °) 10 to 65 Hz 65 to 500 Hz 500 Hz to 1 kHz 1 to 5 kHz 5 to 10 kHz 10 to 30 kHz 0.10 ° 0.25 ° 0.5 ° 2.5 ° 5 ° 10 ° Note See Power and...

[Page 41] Fluke 5522A Operators

Introduction and Specifications Additional Specifications 1 1-21 VARs When the Power Factor approaches 0.0, the Watts output uncertainty becomes unrealistic because the dominant characteristic is the VARs (volts-amps-reactive) output. In the...

[Page 42] Fluke 5522A Operators

5522A Operators Manual 1-22 in this example, the specification is 0.015 % +4 mV as the 0.015 % is the same, and the floor is twice the value (2 x 2 mV). AUX (50th Harmonic) Output: 100 mV, 5 kHz ...................................................

[Page 43] Fluke 5522A Operators

Introduction and Specifications Additional Specifications 1 1-23 AC Voltage (Non-Sine Wave) (cont.) Triangle Wave & Truncated Sine Range, p-p [1] Frequency 1-Year Absolute Uncertainty, tcal ±5 °C, ±(% of output + % of range) [2...

[Page 44] Fluke 5522A Operators

5522A Operators Manual 1-24 AC Voltage, DC Offset Range [1] (Normal Channel) Offset Range [2] Max Peak Signal 1-Year Absolute Uncertainty, tcal ±5 °C [3] ±(% dc output + floor) Sine Waves (rms) 3.3 to 32.999 mV 0 to 50 mV 80 mV...

[Page 45] Fluke 5522A Operators

Introduction and Specifications Additional Specifications 1 1-25 AC Current (Non-Sine Wave) (cont.) Triangle Wave & Truncated Sine Wave Range p-p Frequency 1-Year Absolute Uncertainty tcal ±5 °C ±(% of output + % of range) Max ...

[Page 46] Fluke 5522A Operators

5522A Operators Manual 1-26 ...

[Page 47] Fluke 5522A Operators

2-1 Chapter 2 Preparing for Operations Title Page Introduction.......................................................................................................... 2-3 Unpack and Inspect........................................................

[Page 48] Fluke 5522A Operators

5522A Operators Manual 2-2 ...

[Page 49] Fluke 5522A Operators

2-3 Introduction This chapter provides instructions for unpacking and installing the Calibrator, selecting the line voltage, replacing the fuse, and connecting to line power. Instructions for cable connections other than line power can be found i...

[Page 50] Fluke 5522A Operators

5522A Operators Manual 2-4 How to Select Line Voltage The calibrator arrives from the factory configured for the line voltage normally appropriate for the country of purchase, or as specified at the time of your purchase order. You can operate ...

[Page 51] Fluke 5522A Operators

Preparing for Operations How to Select Line Frequency 2 2-5 100V/120V 220V/240V T5.0A 250V (SB) T2.5A 250V (SB) /63Hz 47Hz 600VA MAX WARNING: TO AVOID ELECTRIC SHOCK GROUNDING CONNECTOR IN POWER CORD MUST BE CONNEC...

[Page 52] Fluke 5522A Operators

5522A Operators Manual 2-6 LC-1 LC-2 LC-3 LC-4 LC-5 LC-6 LC-7 nn008f.eps Figure 2-2. Line Power Cord Types Available from Fluke How to Contact Fluke To order accessories, receive operating assistance, or get the location of the nearest Fluke ...

[Page 53] Fluke 5522A Operators

Preparing for Operations Placement 2 2-7 Placement  Warning To prevent possible electrical shock, fire, or personal injury, make sure that the Product is grounded before use. You may place the calibrator on a bench top or mount it i...

[Page 54] Fluke 5522A Operators

5522A Operators Manual 2-8 ...

[Page 55] Fluke 5522A Operators

3-1 Chapter 3 Features Title Page Introduction.......................................................................................................... 3-3 Front-Panel Features .....................................................................

[Page 56] Fluke 5522A Operators

5522A Operators Manual 3-2 ...

[Page 57] Fluke 5522A Operators

3-3 Introduction This chapter is a reference for the functions and locations of the 5522A Calibrator's front and rear-panel features. Please read this information before operating the calibrator. Front panel operating instructions for the ca...

[Page 58] Fluke 5522A Operators

5522A Operators Manual 3-4 POWER 0 • 1 23 456 7 8 9 ENTER M k m V Hz FIELD EDIT / + F OPR EARTH SCOPEEXGRD MENU PREV SHIFT RESET CE SETUP REF NEW TC MEAS F n p p W dBm sec C A MULT x DIV MODES MORE STBY 5522A CALIBRATOR 3 4 5 6 7 8 9 1 2 gjh00...

[Page 59] Fluke 5522A Operators

Features Softkey Menu Trees 3 3-5 in the lower left corner of the Output Display. In standby mode, the NORMAL, AUX and 20A output terminals are internally disconnected from the 5522A. The 5522A starts up in standby mode. The 5522A automatic...

[Page 60] Fluke 5522A Operators

5522A Operators Manual 3-6 POWER 0 • 1 23 456 7 8 9 ENTER M k m V Hz FIELD EDIT / + F OPR EARTH SCOPEEXGRD MENU PREV SHIFT RESET CE SETUP REF NEW TC MEAS F n p p W dBm sec C A MULT x DIV MODES MORE STBY 5522A CALIBRATOR 10 11 12 13 14 151718 21 ...

[Page 61] Fluke 5522A Operators

Features Softkey Menu Trees 3 3-7  The power switch turns the power on and off. The switch is a latching push-push type. When the switch is latched in, power is on.  m The MORE MODES key provides access to the measure pressure fu...

[Page 62] Fluke 5522A Operators

5522A Operators Manual 3-8 POWER 0 • 1 23 456 7 8 9 ENTER M k m V Hz FIELD EDIT / + F OPR EARTH SCOPEEXGRD MENU PREV SHIFT RESET CE SETUP REF NEW TC MEAS F n p p W dBm sec C A MULT x DIV MODES MORE STBY 5522A CALIBRATOR 32 30 31 29 27 28 26 25 2...

[Page 63] Fluke 5522A Operators

Features Softkey Menu Trees 3 3-9 Table 3-1. Front-Panel Features (cont.)  The TC (Thermocouple) minijack is used for thermocouple simulation during thermometer calibration, and thermocouple measurements. You must use the correct thermo...

[Page 64] Fluke 5522A Operators

5522A Operators Manual 3-10 CALIBRATION ENABLE SERIAL 1 FROM HOST IEEE-488 IN OUT /63Hz 47Hz 600VA MAX FLUKE CORPORATION EVERETT WA, USA NORMAL WARNING: TO AVOID ELECTRIC SHOCK GROUNDING CONNECTOR IN POW...

[Page 65] Fluke 5522A Operators

Features Softkey Menu Trees 3 3-11 Table 3-2. Rear-Panel Features (cont.)   Warning To avoid shock hazard, connect the factory supplied three- conductor line power cord to a properly grounded power outlet. Do not use a two-conduc...

[Page 66] Fluke 5522A Operators

5522A Operators Manual 3-12 SETUP X W AC GI B F C AD G E D AA AB O K H I L P M RQ N S T V U Y Z S2 S3 A Front Panel Key G2 AF AE AJ AG AH AK AL AF Next Section AH AF S1 gjh006.eps Figure 3-3. Setup Softkey Menu Tree ...

[Page 67] Fluke 5522A Operators

Features Softkey Menu Trees 3 3-13 A SHOW SPECS is an online summary of the programmed output specifications. to X W to G to B to F to C to E to D If self test does not pass, error codes are displayed. (See chapter 7, "Maintenance"...

[Page 68] Fluke 5522A Operators

5522A Operators Manual 3-14 E Actual revision numbers replace the numbers in each of the above. Format NV (non-volatile) Memory should be used with caution. Changes are non-reversible. The softkeys function only when the rear-panel CALIBRATION sw...

[Page 69] Fluke 5522A Operators

Features Softkey Menu Trees 3 3-15 I STALL refers to the method of controlling data flow: software control (xon/off), hardware control (rts/cts) or none. STALL refers to the method of controlling data flow: software control (xon/off), hardwar...

[Page 70] Fluke 5522A Operators

5522A Operators Manual 3-16 to N to K EOF (End of File) indicates the action taken at the end of a file by entering one or two ASCII characters. EOF (End of File) ASCII characters are entered with a range of 000 to 255 (first character) and 000 to...

[Page 71] Fluke 5522A Operators

Features Softkey Menu Trees 3 3-17 O GPIB (General Purpose Interface Bus) selects the port address when using the IEEE-488 bus. The factory default is 4. to R to Q DISPLAY BRIGHTNESS and DISPLAY CONTRAST apply to both the Output Display and C...

[Page 72] Fluke 5522A Operators

5522A Operators Manual 3-18 S1 S2 to S3 S to S2 to Tto S1 gjh032.eps Figure 3-4. SETUP Softkey Menu Displays (cont.) ...

[Page 73] Fluke 5522A Operators

Features Softkey Menu Trees 3 3-19 V to AA W X to ACto Y SHOW SPECS is an online summary of the programmed output specifications. U S3 to V to U The values set here become the new limits and can be changed only with new entries or returned t...

[Page 74] Fluke 5522A Operators

5522A Operators Manual 3-20 AA 5522A CAL opens the calibration menu. Refer to the Service Manual for instructions. ZERO zeros the 5522A Calibrator. OHMS ZERO zeros the ohms portion of the 5522A Calibrator; ERR ACT (Error Action) set backup, abor...

[Page 75] Fluke 5522A Operators

Features Softkey Menu Trees 3 3-21 AF AG to AH to AJ to AG AH AE to AF (Only if scope option installed) GO ON and ABORT softkeys are used in the 5522A Calibrator calibration procedure. See the Service Manual for more information. AD to AF g...

[Page 76] Fluke 5522A Operators

5522A Operators Manual 3-22 Table 3-3. Factory Defaults for SETUP Menus Power-Up Defaults Parameter Setting SETUP Menu (Figure 3-4.) User report string (*PUD string) Cleared. D Error units > 0.1% G1 SC-600 option overload test safety...

[Page 77] Fluke 5522A Operators

4-1 Chapter 4 Front Panel Operation Title Page Introduction.......................................................................................................... 4-3 How to Turn on the Calibrator................................................

[Page 78] Fluke 5522A Operators

5522A Operators Manual 4-2 Sine Wave........................................................................................................ 4-41 Triangle Waves........................................................................................

[Page 79] Fluke 5522A Operators

4-3 Introduction  Warning The Calibrator is capable of supplying lethal voltages. To avoid shock hazard, do not make connections to the output terminals when any voltage is present. Placing the instrument in standby may not be enough to a...

[Page 80] Fluke 5522A Operators

5522A Operators Manual 4-4 nn062f.eps After self-test, the control display shows the reset condition (below). nn063f.eps For a discussion of the softkey selection shown above (auto/locked), see “Auto Range Versus Locked Range” late...

[Page 81] Fluke 5522A Operators

Front Panel Operation How to Use the Setup Menu 4 4-5 nn064f.eps This is the primary instrument setup menu. The list below describes submenus available through each softkey and tells you where you can find further information in the manua...

[Page 82] Fluke 5522A Operators

5522A Operators Manual 4-6 Utility Functions Menu The Setup Menu softkey labeled UTILITY FUNCTNS (Utility Functions) provides access to Self Test, Format Nonvolatile Memory, and Instrument Configuration. nn066f.eps • SELF TEST This softk...

[Page 83] Fluke 5522A Operators

Front Panel Operation How to Reset the Calibrator 4 4-7 How to Reset the Calibrator At any time during front panel operation (not remote operation), you can return the Calibrator to the power-up state by pressing R, except after an error m...

[Page 84] Fluke 5522A Operators

5522A Operators Manual 4-8 Operate and Standby Modes When the OPERATE annunciator is lit and OPR is displayed, the output value and function shown on the Output Display is active at the selected terminals. When STBY is displayed in the Output D...

[Page 85] Fluke 5522A Operators

Front Panel Operation How to Connect the Calibrator to a UUT 4 4-9 Recommended Cable and Connector Types  Warning To prevent possible electrical shock, fire, or personal injury, do not touch exposed metal on banana plugs, they can ha...

[Page 86] Fluke 5522A Operators

5522A Operators Manual 4-10 sourced output, a softkey LOs appears, which allows you to tie or open an internal connection between the NORMAL LO terminal and AUX LO terminal. When tied and Z is on, then both LO terminals are tied to chassis groun...

[Page 87] Fluke 5522A Operators

Front Panel Operation How to Connect the Calibrator to a UUT 4 4-11 When calibrating thermocouples, it is especially important to use the correct hookup wire and miniconnector between the Calibrator front panel TC terminal and the UUT. You ...

[Page 88] Fluke 5522A Operators

5522A Operators Manual 4-12 5522A CALIBRATOR UUT HI LO A HI LO INPUT SENSE 4-WIRE W 5522A UUT SENSE SENSE SOURCE SOURCE gjh014.eps Figure 4-2. UUT Connection: Resistance (Four-Wire Compensation) 5522A CALIBRATOR 5522A UUT 87 TRUE RMS MUL...

[Page 89] Fluke 5522A Operators

Front Panel Operation How to Connect the Calibrator to a UUT 4 4-13 5522A CALIBRATOR UUT 5522A 87 TRUE RMS MULTIMETER MIN MAX RANGE HOLD H REL Hz mA A A m mV V V OFF PEAK MIN MAX A COM V mA A gjh016.eps Figure 4-4. UUT Connection: Res...

[Page 90] Fluke 5522A Operators

5522A Operators Manual 4-14 5522A CALIBRATOR 87 TRUE RMS MULTIMETER MIN MAX RANGE HOLD H REL Hz mA A A m mV V V OFF PEAK MIN MAX A COM V mA A gjh018.eps Figure 4-6. UUT Connection: Capacitance (Compensation Off) 5522A CALIBRATOR 87 TRUE R...

[Page 91] Fluke 5522A Operators

Front Panel Operation How to Connect the Calibrator to a UUT 4 4-15 5522A CALIBRATOR 87 TRUE RMS MULTIMETER MIN MAX RANGE HOLD H REL Hz mA A A m mV V V OFF A COM V mA A PEAK MIN MAX gjh020.eps Figure 4-8. UUT Connection: DC Current/AC...

[Page 92] Fluke 5522A Operators

5522A Operators Manual 4-16 5522A CALIBRATOR ! 60V 24V MAX OFFSET ON/OFF F/C HOLD K/J THERMOMETER 51 Connection wiring must match thermocouple type, e.g., K, J, etc. gjh022.eps Figure 4-10. UUT Connection: Temperature (Thermocouple) RMS Versus...

[Page 93] Fluke 5522A Operators

Front Panel Operation Auto Range Versus Locked Range 4 4-17 Auto Range Versus Locked Range A softkey is provided to toggle between the ranging method auto or locked. This feature is available only for single-output dc volts and dc current o...

[Page 94] Fluke 5522A Operators

5522A Operators Manual 4-18 How to Set DC Voltage Output Complete the following procedure to set a dc voltage output at the front panel NORMAL terminals. If you make an entry error, press Gto clear the display, then reenter the value.  Caut...

[Page 95] Fluke 5522A Operators

Front Panel Operation How to Set Output 4 4-19 nn063f.eps • Range (Operating Range) selects autorange (auto) or lock (locked) for the present range. When auto (the default setting) is selected, the calibrator automatically selects th...

[Page 96] Fluke 5522A Operators

5522A Operators Manual 4-20 nn227f.eps Note At voltage outputs of 100 V and above (nominal), you may notice a slight high-pitched sound. This is normal. 1. Press a multiplier key, if necessary. For example, press c. 2. Output in volts. Pr...

[Page 97] Fluke 5522A Operators

Front Panel Operation How to Set Output 4 4-21 nn075f.eps 6. Press O to activate the calibrator output. Several softkey labels appear on the Control Display in the ac voltage function, depending on which waveform is selected: DUTY, OFFS...

[Page 98] Fluke 5522A Operators

5522A Operators Manual 4-22 How to Set DC Current Output Complete the following procedure to set a dc current output between AUX HI and LO or AUX 20A and LO, depending on the current level selected. Current greater than ±3 A is sourced between...

[Page 99] Fluke 5522A Operators

Front Panel Operation How to Set Output 4 4-23 How to Set AC Current Output Complete the following procedure to set an ac current output at the AUX or 20A terminals. If you make an entry error, press G to clear the display, then reenter the...

[Page 100] Fluke 5522A Operators

5522A Operators Manual 4-24 nn321f.eps • Φ & REF MENUS (Phase Difference and 10 MHz reference source.) Selects the phase difference between the NORMAL and AUX outputs, selects internal or external 10 MHz reference, and sets the phas...