[Page 1] Oakton ION 5 Acorn Series
DISTRIBUTED BY: 00702-79 OPERATING INSTRUCTIONS OAKTON WD-35613-21, -30, -31, -33 ION 5 and ION 6 Acorn Series Meters Printed in the U.S.A. 8/99 ION 6 Acorn series pH/Ion/°C Meter ON OFF CAL HOLD ENTER MODE 10. Warranty OAKTON warrants this meter...
[Page 2] Oakton ION 5 Acorn Series
32 Table of Contents 1. Introduction ............................................................................................4 2. Display and keypad functions.................................................................5 3. Preparation..........
[Page 3] Oakton ION 5 Acorn Series
54 1. Introduction Thank you for purchasing an OAKTON ® ION Acorn Meter. These economical, microprocessor-based meters measure the ion concentration of various ions (mono and divalent). The Acorn ION meters feature: • Large LCD for clear and easy...
[Page 4] Oakton ION 5 Acorn Series
3. Preparation Inserting the batteries The battery compartment is found at the back of the instrument. To open the battery compartment, push the lid up. See figure Note the polarity of battery before insert- ing the batteries into position. After r...
[Page 5] Oakton ION 5 Acorn Series
6. If you DO wish to calibrate to this point, press the ENTER key. The meter will display the mV output of your electrode in the standard solution. Allow the instrument to stabilize. Stabilization may take a few minutes depending on electrode and sta...
[Page 6] Oakton ION 5 Acorn Series
1110 pH buffer set selection (ION 6 meter only) The ION 6 meter is capable of up to three-point pH calibration to ensure accuracy across the entire range of the meter. The ION 6 meter can be set to recognize either of two buffer sets: USAor NIST buff...
[Page 7] Oakton ION 5 Acorn Series
NOTE: This meter has automatic buffer recogni- tion which identifies the correct pH buffer values during calibration. If buffers other than 4.01, 7.00 or 10.00 are used, or the electrode has worn out, the LCD will flash "Er1". See figure Pr...
[Page 8] Oakton ION 5 Acorn Series
1514 5. Measurement Measuring Ion Concentration 1. Switch on meter. The ION 5 and ION 6 meters will automatically enter the ion measurement mode. No annunciator will appear. See figure 2. Prepare samples as necessary (i.e., add Ionic Strength adjust...
[Page 9] Oakton ION 5 Acorn Series
1716 6. Electrode care and maintenance Ion selective electrode care For Ion Selective Electrode care, consult the instruction manual included with your electrode. pH electrode care (ION 6 meter only) Storage For best results, keep the pH bulb wet. ...
[Page 10] Oakton ION 5 Acorn Series
1918 8. Specifications Mode Ion Temperature mV pH (ION 6 only) Range 0.1 to 1999 ppm 0.00 to -500 to 0.0 to 100°C +500 mV 14.00 pH 0.1 ppm from 0.1 mV, Resolution 0.1 ppm-199.9 ppm 0.1°C ±199.9 mV; 0.01 pH 1 ppm from 1 mV outside 200-1999 ppm...
[Page 11] Oakton ION 5 Acorn Series
2120 WD-35812-42 Sodium ion selective electrode, includes 15 ml of reference electrolyte and a filling pipette. Shpg wt 0.5 lb (230 g). WD-35812-93 Sodium ion selective solution kit, includes 475 ml Ion Strength Adjuster, 475 ml calibration standard,...