Olympus CX22LED Instructions Manual

Download or browse on-line these Instructions Manual for Olympus CX22LED Microscope.

Olympus CX22LED Manual Information:

This manual for Olympus CX22LED, given in the PDF format, is available for free online viewing and download without logging on. The guide contains 28 pages, and the size of the file at download is . The document type is Instructions Manual.

Download Manual

Summary of Contents:

[Page 1] Olympus CX22LED

INSTRUCTIONS CX22LED BIOLOGICAL MICROSCOPE A X 8 0 0 2 This instruction manual is for the Olympus Biological Microscopes Model CX22LED. To ensure the safety, obtain optimum performance and to familiarize yourself fully with the use of this microscope...

[Page 2] Olympus CX22LED

NOTE: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is...

[Page 3] Olympus CX22LED

CX22LED CONTENTS 4 5 6 7-11 IMPORTANT — Be sure to read this section for safe use of the equipment. — 1-3 10-1 System Diagram of Optional Accessories ............................................................................ 18 10-2 Install...

[Page 4] Olympus CX22LED

1 IMPORTANT SAFETY PRECAUTIONS 1. After the equipment has been used in an observation of a specimen that is accompanied with a potential of infection, clean the parts coming in contact with the specimen to prevent infection. · Moving this product i...

[Page 5] Olympus CX22LED

2 CX22LED 1 Intended use This product has been designed to be used to observe magnified images of specimens in various routine work and research applications. Do not use this product for any purpose other than its intended use. Disconnect the AC ...

[Page 6] Olympus CX22LED

3 3 Maintenance and Storage 1. To clean the lenses and other glass components, simply blow away dirt using a commercially available blower and wipe the components gently using a piece of cleaning paper (or clean gauze). If a lens is stained with ...

[Page 7] Olympus CX22LED

4 CX22LED STANDARD COMPONENT UNITS }After opening the package, make sure that the correct component units for the selected set are included. The 100X objective and immersion oil are provided only with the CX22LEDRFS1 set. The objectives have been scr...

[Page 8] Olympus CX22LED

5 NOMENCLATURE }The following items have been attached at the factory to prevent deterioration during transport. Remove these items and retain them for future use. Then loosen the observation tube clamping knob and correct the orientation of the eyep...

[Page 9] Olympus CX22LED

6 CX22LED Start observation. SUMMARY OF BRIGHTFIELD OBSERVATION PROCEDURE 1Main switch (P. 7) 2Light intensity adjustment knob (P. 7) 3Specimen holder (P. 7) 4X-axis/Y-axis feed knobs (P. 7) 5Revolving nosepiece (P. 10) 6Coarse/fine focus adjustment ...

[Page 10] Olympus CX22LED

7 Fig. 5 Fig. 4 DETAILED OBSERVATION PROCEDURE 1 Turning the LED Illumination ON (Fig. 4) 1. Set the main switch 1 to “ I ” (ON). 2. Rotating the light intensity adjustment knob 2 in the direction of the arrow increases brightness and rotatin...

[Page 11] Olympus CX22LED

8 CX22LED Fig. 7 Specimen holder scales (Fig. 7) }Theses scales allow the position (coordinates) being observed on the specimen to be identified. Even after the specimen is moved, it can be returned easily to the original position. 1. The horizontal ...

[Page 12] Olympus CX22LED

9 Fig. 10 Pre-focusing Knob (Fig. 10) }The pre-focusing knob controls the mechanism for preventing collision between the specimen and objective. 1. After bringing the specimen into focus, turn the pre-focusing knob 1 inside the hole of the arm so tha...

[Page 13] Olympus CX22LED

10 CX22LED Fig. 14 6 Adjusting the Condenser Position and Aperture Iris Diaphragm (Fig. 14) }The condenser is usually used in the highest position. If the entire observed field of view is not bright enough, brightness may be improved by lowering ...

[Page 14] Olympus CX22LED

11 Fig. 16 8 Using the 100X Immersion Objective (Fig. 16) }The designated immersion oil should be applied to the top lens of the 100X immersion objective. Otherwise, the observed image will be unable to be focused on. Always use the immersion oil...

[Page 15] Olympus CX22LED

12 CX22LED ONE-POINT ADVICE How To Track a Microscopic Image To track an image moving leftwards (or to move the specimen image rightwards): Direction of moving the image Total Magnification The size of the specimen image for observation is obtained b...

[Page 16] Olympus CX22LED

13 TROUBLESHOOTING GUIDE Under certain conditions, performance of the unit may be adversely affected by factors other than defects. If problems occur, please review the following list and take remedial action as needed. If you cannot solve the proble...

[Page 17] Olympus CX22LED

14 CX22LED Problem Cause Remedy Page 7 7 17 12. Objective hits the specimen when an objective is switched to a higher- magnification objective. The specimen is upside down. Set the specimen correctly with the cover glass on the top. The cover glass i...

[Page 18] Olympus CX22LED

15 SPECIFICATIONS Item Specifications 1. Optical system Infinity Correction optical system 2. Illumination Built-in 0.5 W LED illumination system. Rated voltage: 5-6 V 0.5 A (AC adapter: 100-240 V 50/60 Hz 0.4A) Power cons...

[Page 19] Olympus CX22LED

16 CX22LED OPTICAL CHARACTERISTICS Optical Characteristics Objectives Magnifi- cation N.A. W.D. (mm) Cover Glass Thickness (mm) Resolution (μm) 10X Eyepieces (FN 20) Total Power Focal Depth (μm) Actual Field of View Remark Plan objective (FN 20) ...

[Page 20] Olympus CX22LED

17 Fig. 18 1 · Be sure to use the provided AC adapter. Use of any other AC adapter may cause a malfunction. · The power cord and AC adapter cord are vulnerable when bent or twisted. Never subject them to excessive force. · Make sure that the ma...

[Page 21] Olympus CX22LED

18 CX22LED Fig. 19 1 Filter Holder CH2-FH and Darkfield Ring CH2-DS (Fig. 19) }This holder accommodates the CH2-DS darkfield ring. 1. Insert the darkfield ring @ in the CH2-FH filter holder 2. 2. Push the filter holder 2 containing the darkfield ...

[Page 22] Olympus CX22LED

19 2 Reflection Mirror CH20-MM (Fig. 20) }The reflection mirror is designed to be used in microscopic observation in a location where power supply is not available. It makes it possible to use the natural light in place of the LED illumination. F...

[Page 23] Olympus CX22LED

20 CX22LED }Prepare one eyepiece micrometer disk (diameter 20.4 mm, thickness 1 mm) and two 20.4-RH reticle holders (available as 2-piece set). The field number becomes 19.6 when the reticle holders are used. If you are using WHC15X or WHC15X-H, plea...

[Page 24] Olympus CX22LED

21 Fig. 25 Fig. 26 Fig. 27 2. Place the AC adapter 3 on top a of the cord hanger 4 (Fig. 25). 3. Bring the cable 5 of the AC adapter into the hole of center part b of cord hanger. Then, bring it down so that it goes around the bottom part c...

[Page 25] Olympus CX22LED

22 CX22LED · Do not carry the microscope by holding it by the cord hanger. If it is detached during transport, the microscope may drop and human injury may result. · Be sure to disconnect the AC adapter from the microscope when storing the micros...

[Page 26] Olympus CX22LED

23 PROPER SELECTION OF THE POWER CORD If no power cord is provided, please select the proper power cord for the equipment by referring to “ Specifications ” and “ Certified Cord ” below: CAUTION: In case you use a non-approved power cord...

[Page 27] Olympus CX22LED

24 CX22LED Table 2 HAR Flexible Cord APPROVAL ORGANIZATIONS AND CORDAGE HARMONIZATION MARKING METHODS Approval Organization Printed or Embossed Harmoniza- tion Marking (May be located on jacket or insulation of internal wir- ing) Alternative Marking...

[Page 28] Olympus CX22LED

Shinjuku Monolith, 3-1, Nishi Shinjuku 2-chome, Shinjuku-ku, Tokyo, Japan Wendenstrasse 14-18, 20097 Hamburg, Germany 3500 Corporate Parkway, Center Valley, Pennsylvania 18034-0610, U.S.A. 491B River Valley Road, #12-01/04 Valley Point Office Tower, ...