[Page 1] Rigol DP832

RIGOL User’s Guide DP800 Series Programmable Linear DC Power Supply Jun. 2016 RIGOL TECHNOLOGIES, INC. ...

[Page 2] Rigol DP832

...

[Page 3] Rigol DP832

RIGOL DP800 User’s Guide I Guaranty and Declaration Copyright © 2012 RIGOL TECHNOLOGIES, INC. All Rights Reserved. Trademark Information RIGOL is a registered trademark of RIGOL TECHNOLOGIES, INC. Publication Number UGH04109-1110 ...

[Page 4] Rigol DP832

RIGOL II DP800 User’s Guide Safety Requirement General Safety Summary Please review the following safety precautions carefully before putting the instrument into operation so as to avoid any personal injury or damage to the instrument...

[Page 5] Rigol DP832

RIGOL DP800 User’s Guide III Do Not Operate With Suspected Failures. If you suspect that any damage may occur to the instrument, have it inspected by RIGOL authorized personnel before further operations. Any maintenance, adjustment or repla...

[Page 6] Rigol DP832

RIGOL IV DP800 User’s Guide Safety Notices and Symbols Safety Notices in this Manual: WARNING Indicates a potentially hazardous situation or practice which, if not avoided, will result in serious injury or death. CAUTION ...

[Page 7] Rigol DP832

RIGOL DP800 User’s Guide V Allgemeine Sicherheits Informationen Überprüfen Sie diefolgenden Sicherheitshinweise sorgfältigumPersonenschädenoderSchäden am Gerätundan damit verbundenen weiteren Gerätenzu vermeiden. Zur Vermeidung von...

[Page 8] Rigol DP832

RIGOL VI DP800 User’s Guide Betreiben Sie das Gerät nicht im Fehlerfall. Wenn Sie am Gerät einen Defekt vermuten, sorgen Sie dafür, bevor Sie das Gerät wieder betreiben, dass eine Untersuchung durch RIGOL autorisiertem Personal durchge...

[Page 9] Rigol DP832

RIGOL DP800 User’s Guide VII Sicherheits Begriffe und Symbole Begriffe in diesem Guide: WARNING Die Kennzeichnung WARNING beschreibt Gefahrenquellen die leibliche Schäden oder den Tod von Personen zur Folge haben können. CAUT...

[Page 10] Rigol DP832

RIGOL VIII DP800 User’s Guide Care and Cleaning Care Do not store or leave the instrument where it may be exposed to direct sunlight for long periods of time. Cleaning Clean the instrument regularly according to its operating conditi...

[Page 11] Rigol DP832

RIGOL DP800 User’s Guide IX Environmental Considerations The following symbol indicates that this product complies with the requirements in WEEE Directive 2002/96/EC. Product End-of-Life Handling The equipment may contain substan...

[Page 12] Rigol DP832

RIGOL X DP800 User’s Guide ...

[Page 13] Rigol DP832

RIGOL DP800 User’s Guide XI DP800 Series Overview DP800 series is high-performance programmable linear DC power supply. DP800 series which provides clear user interface, superb performance specifications, various analysis functions as wel...

[Page 14] Rigol DP832

RIGOL XII DP800 User’s Guide cycles of channel on/off toggle  Support digital trigger (option) to realize digital trigger input and trigger output functions  Support on-line analyzer (option) to perform on-line analysis of variou...

[Page 15] Rigol DP832

RIGOL DP800 User’s Guide XIII Document Overview Chapter 1 Quick Start Introduce the appearance and dimensions, front panel, rear panel as well as user interface of DP800. In addition, it provides the detailed procedures of power connecti...

[Page 16] Rigol DP832

RIGOL XIV DP800 User’s Guide Format Conventions in this Manual 1. Key The function key at the front panel is denoted by the format of "Key Name (Bold) + Text Box" in the manual. For example, Utility denotes the "Utility&q...

[Page 17] Rigol DP832

Contents RIGOL DP800 User’s Guide XV Contents Guaranty and Declaration ......................................................................... I Safety Requirement .............................................................................

[Page 18] Rigol DP832

RIGOL Contents XVI DP800 User’s Guide Display Setting ..................................................................................... 2-41 Brightness .................................................................................... 2...

[Page 19] Rigol DP832

Chapter 1 Quick Start RIGOL DP800 User’s Guide 1-1 Chapter 1 Quick Start The contents of this chapter are as follows:  General Inspection  Appearance and Dimensions  Front Panel  Rear Panel  To Connect to Powe...

[Page 20] Rigol DP832

RIGOL Chapter 1 Quick Start 1-2 DP800 User’s Guide General Inspection 1. Inspect the packaging If the packaging has been damaged, do not dispose the damaged packaging or cushioning materials until the shipment has been checked for comple...

[Page 21] Rigol DP832

Chapter 1 Quick Start RIGOL DP800 User’s Guide 1-3 Appearance and Dimensions Figure 1-1 Front View Unit: mm Figure 1-2 Side View Unit: mm ...

[Page 22] Rigol DP832

RIGOL Chapter 1 Quick Start 1-4 DP800 User’s Guide Front Panel This section introduces the front panel of DP800 series by taking DP832 (as shown in the figure below) as an example. The differences of different models are introduced separa...

[Page 23] Rigol DP832

Chapter 1 Quick Start RIGOL DP800 User’s Guide 1-5 1. LCD 3.5 inches TFT display. It is used to display the system parameter setting, system output state, menu items, prompt messages, etc. 2. Channel (Range) Selection and Output Switch ...

[Page 24] Rigol DP832

RIGOL Chapter 1 Quick Start 1-6 DP800 User’s Guide Single-channel model (DP811): 20V 10A 40V 5A Press this key to select 20V/10A range as the current range and you can set the parameters of the channel, such as voltage, current a...

[Page 25] Rigol DP832

Chapter 1 Quick Start RIGOL DP800 User’s Guide 1-7 the keys are used to enter the voltage units (V and mV) or the current units (A and mA). (2) Numeric Keyboard Ring-type numeric keyboard: include numbers 0-9 and the decimal point. Press...

[Page 26] Rigol DP832

RIGOL Chapter 1 Quick Start 1-8 DP800 User’s Guide 7. Output Terminals The output terminals of different models of DP800 series are different. DP832: DP821: 8V/10A60V/1A DP811: S S Output (1) Channel outp...

[Page 27] Rigol DP832

Chapter 1 Quick Start RIGOL DP800 User’s Guide 1-9 Connection methods of the output terminal: Method 1: Connect the test lead to A of the output terminal. Method 2: Rotate the outer nut of the output terminal counterclockwise and con...

[Page 28] Rigol DP832

RIGOL Chapter 1 Quick Start 1-10 DP800 User’s Guide Press this key to enter the system utility function setting interface. Users can set the remote interface parameters, system parameters and print parameters. Besides, users can also cali...

[Page 29] Rigol DP832

Chapter 1 Quick Start RIGOL DP800 User’s Guide 1-11 Rear Panel This section introduces the rear panel of DP800 series by taking DP832 and DP811 (as shown in the figures below) as examples. The introduction of each part is as shown in Table...

[Page 30] Rigol DP832

RIGOL Chapter 1 Quick Start 1-12 DP800 User’s Guide Table 1-1 DP800 rear panel explanation No. Name Explanation 1 LAN Interface (option) The instrument is connected to the local network via RJ45 interface 2 USB DEVICE Connect t...

[Page 31] Rigol DP832

Chapter 1 Quick Start RIGOL DP800 User’s Guide 1-13 To Connect to Power DP800 series power supply supports various AC power supply inputs. The voltage selector setting at the rear panel differs when the input power connected is different, ...

[Page 32] Rigol DP832

RIGOL Chapter 1 Quick Start 1-14 DP800 User’s Guide Power-on Inspection Press the power switch at the front panel, the instrument starts and executes self-test. If the instrument passes the self-test, the welcome interface will be display...

[Page 33] Rigol DP832

Chapter 1 Quick Start RIGOL DP800 User’s Guide 1-15 3. If needed, adjust the power voltage selector manually to select the voltage scale (please refer to Table 1-4) that matches the actual input voltage. 4. Take out the fuse and replace ...

[Page 34] Rigol DP832

RIGOL Chapter 1 Quick Start 1-16 DP800 User’s Guide User Interface DP800 series power supply provides three kinds of display modes (normal, waveform and dial). The default is normal. You can press Display  Disp Mode to select different...

[Page 35] Rigol DP832

Chapter 1 Quick Start RIGOL DP800 User’s Guide 1-17 Table 1-4 User Interface Explanation No. Explanation 1 Channel number 2 Channel output voltage/current 3 Channel output status 4 Channel output mode 5 Status bar. Display the...

[Page 36] Rigol DP832

RIGOL Chapter 1 Quick Start 1-18 DP800 User’s Guide To Use the Built-in Help System The built-in help system provides help information for any front panel key (except the parameter input area) and menu keys for users to quickly obtain the ...

[Page 37] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-1 Chapter 2 Front Panel Operations The contents of this chapter are as follows:  Constant Voltage Output  Constant Current Output  Power Supply Series and Parall...

[Page 38] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-2 DP800 User’s Guide Constant Voltage Output DP800 series power supply provides three output modes: constant voltage output (CV), constant current output (CC) and critical mode (UR). In CV mode, the...

[Page 39] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-3 select the desired unit. Besides, you can also press OK to select the default unit (V). During the input, you can press Back to delete the character currently before the cursor or ...

[Page 40] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-4 DP800 User’s Guide Warning To avoid electric shock, please turn on the output switch after the output terminals are correctly connected. CAUTION When the fan stops, the channel switch can ...

[Page 41] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-5 higher output accuracy). Tip In CV output mode, when the load current is greater than the current se...

[Page 42] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-6 DP800 User’s Guide Constant Current Output In constant current output mode, the output current equals the current setting value and the output voltage is determined by the load. This section intro...

[Page 43] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-7 CAUTION When the fan stops, the channel switch cannot be turned on; otherwise, "The fan stops, stop the output!" will be displayed. 8. Check the output mode In co...

[Page 44] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-8 DP800 User’s Guide Power Supply Series and Parallel Connections Higher voltages can be provided when two or more insulated channels (the channels can be from a single power supply or multiple powe...

[Page 45] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-9 2. Set proper voltage, current and overcurrent protection value for each channel (all the channels in series connection should be working in constant voltage mode; you should set ...

[Page 46] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-10 DP800 User’s Guide Track Function The track function is usually used to provide symmetric voltage for the operation amplifier or other circuit. For DP800 series power supply, the specified channe...

[Page 47] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-11 You can perform the following operations according to your need. 1. Only enable the track function of a single channel and track the voltage setting value The operation ...

[Page 48] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-12 DP800 User’s Guide displayed between the CH1 and CH2 areas in the user interface. Method 2  Select the "Independent" tracking mode Press Utility  System  Track Set  Track ...

[Page 49] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-13 Sense Working Mode When the output current of the power supply is high, the voltage drop on the load lead cannot be ignored. To ensure that the load can acquire the correct volt...

[Page 50] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-14 DP800 User’s Guide Sense working mode. The Sense connecting method at the rear panel is as shown in the figure below. Operation Procedures: 1. Connect the load to the corresponding posit...

[Page 51] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-15 Timer and Delayer DP800 provides the timer and delayer functions. When the timer is enabled, the instrument outputs the preset voltage and current values (at most 2048 groups...

[Page 52] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-16 DP800 User’s Guide To Set the Timer Parameters Press Timer  Timer Set to enter the timer parameter setting interface as shown in the figure below. The channel currently selected is display...

[Page 53] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-17 To Set the Number of Output Groups The number of output groups is defined as the number of groups of preset voltage and current values that the power supply outputs in each cycl...

[Page 54] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-18 DP800 User’s Guide To Edit the Timer Parameters Manually You need to set P groups (No.0 to No.(P-1)) of timer parameters; wherein, P is the number of output groups currently set. Only 6 groups of...

[Page 55] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-19  Sine The Sine waveform is as shown in the figure below. The instrument determines the Sine amplitude according to the maximum and minimum currently set and determines the Si...

[Page 56] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-20 DP800 User’s Guide  Ramp The Ramp waveform is as shown in the figure below. The instrument determines the amplitude of the Ramp according to the maximum and minimum currently set, determines t...

[Page 57] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-21  StairUpDn The StairUpDn waveform is as shown in the figure below. The instrument determines the StairUpDn waveform according to the maximum (denoted by MAX), minimum (denote...

[Page 58] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-22 DP800 User’s Guide  Exp Fall The Exp Fall waveform is as shown in the figure below. The instrument determines the Exp Fall waveform according to the maximum (denoted by MAX), minimum (denoted ...

[Page 59] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-23 3. Edit the timer parameters For different templets, the timer parameters to be set are different as shown in Table 2-1. Table 2-1 Timer parameters (templet) Templet Param...

[Page 60] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-24 DP800 User’s Guide  Hi Level When the templet currently selected is Pulse, set the high level of the Pulse and the range is related to the channel currently selected. It should be greater than...

[Page 61] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-25 To Save and Read the Timer File You can store the timer parameters edited manually or using the templet in internal or external memory and recall them when required. 1. Sav...

[Page 62] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-26 DP800 User’s Guide To Set the Delay Parameters Press Timer  Delay Set to enter the delay parameter setting interface as shown in the figure below. The channel currently selected is displayed i...

[Page 63] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-27 To Set the Number of Cycles The number of cycles refers to the number of times that the instrument performs delay output according to the preset state. Press Cycles to set the n...

[Page 64] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-28 DP800 User’s Guide 2. Use the up/down direction key to select the state (State) and time (Delay) of the current group respectively. When state (State) is selected, press OK to switch to the desire...

[Page 65] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-29 3) Set the time base value and step When the time generation method is "Increase" or "Decline", users need to set the time base value and step. Press Time...

[Page 66] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-30 DP800 User’s Guide To Enable the Delayer After setting the delay parameters, return to the timer and delayer interface (as shown in Figure 2-1). Press Delayer to enable the delay output. The dela...

[Page 67] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-31 Advanced Functions DP800 provides various advanced functions, including the recorder, analyzer (option [1] ), monitor (option) and trigger (option). Press to open the advanced...

[Page 68] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-32 DP800 User’s Guide Recorder When the recorder is enabled, users can record the current state of the instrument and if an analyzer is installed, users can also analyze the file recorded. Press...

[Page 69] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-33 Analyzer (option) The analyzer can analyze the file recorded and provides the analysis results of the channel voltage, current and power, such as the average, VAR and range. ...

[Page 70] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-34 DP800 User’s Guide 4. View the analysis results After the analysis is finished, you can view the analysis results according to your need.  Analysis object Press Object to select "Volta...

[Page 71] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-35 Table Display each group of voltage, current and power of the current channel in the record file opened in table form. Tip  You can view the analysis results of the s...

[Page 72] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-36 DP800 User’s Guide Monitor (option) The monitor can monitor the current output state of the instrument. When the user-defined monitor condition is met, the instrument executes the corresponding ...

[Page 73] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-37 Trigger (option) The rear panel of DP800 provides a digital I/O interface which supports trigger input and trigger output.  Trigger Input: The data lines of the digit...

[Page 74] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-38 DP800 User’s Guide Figure 2-6 Trigger Setting Interface Trigger Input When signal that meets the current trigger type is input on the specified data line, the specified source under contro...

[Page 75] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-39 or "Output Toggle".  Output On: when the trigger condition is met, turn on the output of the channel currently selected as the source under control.  Output O...

[Page 76] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-40 DP800 User’s Guide 3. Trigger Condition Press Condition to open the trigger condition setting interface and users can set the trigger condition of trigger output.  Outp Trig: the instrument ...

[Page 77] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-41 Display Setting Press Display at the front panel to enter the interface as shown in the figure below. You can set the screen brightness, contrast, RGB luminance, display mode ...

[Page 78] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-42 DP800 User’s Guide Display Mode Press Display  Disp Mode to set the display mode to "Normal", "Waveform" or "Dial". The default is normal display mode. This setti...

[Page 79] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-43 User-defined Start-up Interface DP800 allows users to define the start-up interface. First, store the content to be displayed in a USB storage device in BMP format (the picture ...

[Page 80] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-44 DP800 User’s Guide 4. Save the settings Press Save  OK to save the start-up interface settings. 5. Select the start-up interface Press Boot UI to select the default start-up interface or t...

[Page 81] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-45 Store and Recall DP800 allows users to store various kinds of files in internal or external memory and recall the files stored when required. DP800 provides an internal non-...

[Page 82] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-46 DP800 User’s Guide Browser Press Store  Browser to move the cursor to the directory or file area and then use the knob or the up/down direction key to select the desired directory or file. ...

[Page 83] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-47 Save You can save a file following the steps below. 1. Press Store  Type to select the desired file type. Note: The save operation is available only when the file type i...

[Page 84] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-48 DP800 User’s Guide  Chinese Input In Chinese input method, use the knob to select the desired Pinyin character in the "Virtual Keyboard" and press Select to select this character; th...

[Page 85] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-49 4. After inputting the filename in the filename input interface, press OK and the instrument stores the file with the specified filename in the specified file type under the stor...

[Page 86] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-50 DP800 User’s Guide Copy and Paste You can only copy the files in C disk and paste the files to D disk. 1. Press Store  Browser to move the cursor to the directory area and select C disk; p...

[Page 87] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-51 Utility Press Utility at the front panel to enter the interface as shown in the figure below. This interface displays the current system parameters. Figure 2-13 Utility I...

[Page 88] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-52 DP800 User’s Guide I/O Configuration DP800 supports USB, LAN (option), RS232 (option) and GPIB (option, extended using the USB to GPIB interface converter) interfaces, via which you can control ...

[Page 89] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-53  DHCP: in this mode, the DHCP server in the current network assigns network parameters (such as the IP address) for the instrument. Press DHCP to enable or disable the DHCP co...

[Page 90] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-54 DP800 User’s Guide VISA descriptor is the resource name and describes the accurate name and location of the VISA resource. If the LAN interface is currently used to communicate with the instrument,...

[Page 91] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-55 Press DNS Serv and use the numeric keyboard and left/right direction key or the knob and left/right direction key to input the desired address. This setting is stored in the non-v...

[Page 92] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-56 DP800 User’s Guide To Set the RS232 Parameters Connect the RS232 interface to the PC or data terminal equipment (DTE) using RS232 cable and set the interface parameters (baud rate, parity bit, et...

[Page 93] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-57 To Set the GPIB Address Before using the GPIB interface, use the USB to GPIB interface converter (option) to extend a GPIB interface and then use a GPIB cable to connect the ins...

[Page 94] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-58 DP800 User’s Guide Screen Saver Press Utility  System  Scrn Svr to enable or disable the screen saver function. When the screen saver function is enabled, the instrument will enter the scre...

[Page 95] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-59 2. Press Utility  System  Track Set  On/Off to select "Enable" or "Disable".  Disable: do not track the output status of the channel.  Enab...

[Page 96] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-60 DP800 User’s Guide Print You can store the screen content as a picture file in the USB storage device. Note: DP800 only supports the flash memory USB storage device of FAT32 format. 1. Inser...

[Page 97] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-61 Preset Setting Press Utility  Preset to define the setting recalled by Preset at the front panel. Pressing Preset can restore the instrument to factory setting or recall the ...

[Page 98] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-62 DP800 User’s Guide Table 2-2 Factory setting Channel Parameters DP832 Voltage/Current Setting Values CH1: 00.00V/3.000A CH2: 00.00V/3.000A CH3: 0.00V/3.000A Voltage/Current Limits CH1: 33.00...

[Page 99] Rigol DP832

Chapter 2 Front Panel Operations RIGOL DP800 User’s Guide 2-63 Display Brightness* 50% Contrast* 25% RGB Luminance* 50% Display Mode* Normal Theme* Green System Setting Language* Chinese(s) Power-on Setting* Default OTP...

[Page 100] Rigol DP832

RIGOL Chapter 2 Front Panel Operations 2-64 DP800 User’s Guide Delayer Channel CH1 Delayer On/off Off Output Groups 1 Delay Parameter State: Off, On alternately Cycles 1 End State Outp Off State Generation 0 1Patt Time Gener...